The Pulse Oximeter and Plethysmograph Sensor measures blood oxygen saturation (SpO2) and pulse pressure (plethysmography) using a dual-wavelength light transmittance based finger clip.
Human pulse oximeter pro outputs the SPO2 and pulse waveform. Measure the Heart Rate using LabScribe software
$1,632.00 – $2,418.00Price range: $1,632.00 through $2,418.00
The Pulse Oximeter measures blood oxygen saturation (SpO2) in humans using a finger or ear clip sensor.
The USB pulse-ox is a stand alone device that comes with LabScribe software. It does not need an additional recorder.
The iWire pulse-ox requires an iWorx recorder with the iWire port, such as the IX-TA, IX-RA or IX-Peak recorders.
The DIN8 pulse-ox required an iWorx Recorder or amplifier with a DIN8 input and a BNC input ( for pulse signal).
| Pulse Oximeter Specifications | |
|---|---|
| O2 Saturation Range: | 80-100% |
| Pulse Rate Range: | 18 to 321 Beats per Minute (BPM) |
| SpO2 range | 80-100% |
| Finger Clip: | +2 digits |
| Ear Clip: | +3 digits |
| Measurement Wavelength | Red – 660 nanometers Infrared – 925 nanometers |
| Temperature range | 0 – +50 degrees C |
| Humidity | 10 – 90% non-condensing |
| Power Draw | 60 mW – typical operating |
| Patient Isolation Dielectric Withstand | Meets IEC 60601-1 |
| Leakage Current | not applicable |
| Integrated Finger sensor Dimension | 1.26 in. x 1.25 in. x 2 in. (32 mm x 32 mm x 51 mm) |
| Box Dimensions | 2.27 in. x 2.4 in. x 1.0 in. 57 mm x 61 mm x 25 mm) |
| Units Conversion | 80 % -> 0V 100% -> 2V ie Slope = 10, offset = 80 |