Search
Generic filters

IX-PeakPro Elite Fitness Assessment Recorder

The IX-Peak is suitable for recording and measuring:

  • Basal Metabolic Rate (BMR)
  • Resting Metabolic Rate (RMR)
  • Respiratory Exchange Ratio (RER)
  • Sedentary to light activity VO2 and VCO2
  • VO2 Max

SKU IX-PREAKPRO Category

The IX-PEAKPRO is part of the MC-PEAKPRO Research Grade Metabolic Cart. 

Specifications
General
sensors Volume, O2, CO2,Temp, Barometric Pressure, HR
iWire 2 iWire ports
Temp 10 – 50 deg C
HR Polar Heart Rate
CO2 Sensor
Range 0-10%
Sense Method NDIR
Linearity < 0.1% CO2
Repeatability < 0.01% CO2
Drift < 0.1% in 1st 24 Hours, < 0.2% the following week.
Warm up 5-8 minutes
O2 Sensor
Range 0-50 %
Sense Method Paramagnetic
Linearity < 0.2% O2
Repeatability < 0.2% O2
Drift < 0.4% in 1st 24 Hours, then <0.2% additional the following week, then < 0.2% additional per each following month.
Warm up 1 minute

Design Your System

Choose the parameters you want to Measure

ECG/EKG
EMG : Number of Channels
Invasive BP : Number of Channels
Non-Invasive BP : Number of Channels
Flow : Number of Channels
Temperature: Number of Channels
Force: Number of Channels
Stimulation: Describe the type