Search
Generic filters

IX-Peak Fitness Assessment Recorder

The IX-Peak is suitable for recording and measuring:

  • Basal Metabolic Rate (BMR)
  • Resting Metabolic Rate (RMR)
  • Respiratory Exchange Ratio (RER)
  • Sedentary to light activity VO2 and VCO2
  • VO2 Max


Manual: IX-PEAK

SKU IX-PEAK Category

The IX-PEAK is part of the MC-PEAK metabolic cart. 

Specifications

General
sensorsVolume, O2, CO2,Temp, Barometric Pressure, HR
iWire2 iWire ports
Temp10 – 50 deg C
HRPolar Heart Rate
CO2 Sensor
Range0-10%
Sense MethodNDIR
Error< 0.26% CO2
Warm up5 minutes
O2 Sensor
Range0-50 %
Sense MethodParamagnetic
Linearity< 0.2% O2
Repeatability< 0.2% O2
Drift< 0.4% in 1st 24 Hours, then <0.2% additional the following week, then < 0.2% additional per each following month.
Warm up1 minute

Design Your System

Choose the parameters you want to Measure

ECG/EKG
EMG : Number of Channels
Invasive BP : Number of Channels
Non-Invasive BP : Number of Channels
Flow : Number of Channels
Temperature: Number of Channels
Force: Number of Channels
Stimulation: Describe the type

Recover Password

We will contact you with your login info. 

Newsletter Signup