Search
Generic filters

IA-400D 4 Channel Amplifier

The IA-400D is a signal conditioner for iWorx-brand transducers that provides four channels of excitation power, gain, and noise filtration. It exhibits low system noise and offset, making it suitable for use with high resolution data acquisition systems.

$1,695.00


Manual: IA-400D

SKU IA-400D Category

The IA-400D is a signal conditioner for iWorx-brand transducers that provides four channels of excitation power, gain, and noise filtration. It exhibits low system noise and offset, making it suitable for use with high resolution data acquisition systems.

Filters

Each channel of the IA-400D has a switch to allow selection of either a 40 Hz or 10 kHz low pass filter.? This makes the IA-400D an ideal solution for conditioning signals from low-bandwidth transducers such as those used to measure blood pressure or force as well as those applications requiring wider bandwidth.

When used with the C-ISO-256 , the IA-400D can be used to measure Biopotentials such as ECG, EMG etc. from humans.

Gain Control

Gain control on the IA-400D is very straightforward. A programming resistor in the DIN-8 connector is used to set the gain. iWorx transducers set the IA-400D gain to the optimal value for a given measurement. However, this does not limit the IA-400D to use with iWorx transducers. Many non-iWorx transducers, can be modified to work with the IA-400D.

Offset

The IA-400D is an inherently low offset device. This, together with the low offsets characteristic of bridge-style transducers and the high resolution of the iWorx data acquisition systems dramatically minimizes the need for offset control. The IA-400D user experience is simplified as a result.

The IA-400D is packaged in a low profile enclosure that provides both durability and noise immunity. The device has a small foot print measuring just 6.2 in. x 5.3 in. x 2.3 in. (17.5 cm x 13.5 cm x 6cm)and multiple units are easily stacked to save bench space.

Citations


Interfacing to other Transducers:

iWorx provides various adaptor cables for connecting your transducer. Take a look at he various standard interface cable options.? For example: The C-DIN-GRASS-F cable can be used to interface with the Grass FT-03 or FT-10 force transducers.

Power Supply: A-UP12-DC-25-1.5A

Specifications

Number of channels Four (DIN8 connectors)
Operational Modes Bridge
Low-pass Filter Switchable 40 Hz and 10K Hz (± 10%) 8 pole Bessel filter
Input Impedance 10 gigOhm
Transducer Excitation Voltage ± 5 VDC
Gain 110K (set by transducer)
CMRR
Gain = 1 > 80 dB (Minimum) / >92 dB (Typical)
Gain = 10 > 95 dB (Minimum) / >110 dB (Typical)
Gain = 100 > 110 dB (Minimum) / >120 dB (Typical)
Gain = 1000 > 112 dB (Minimum) / >135 dB (Typical)
System Noise <10 µV (RTI) / < 2 µV (RTO)
Number of Outputs 4 single-ended BNC connectors
Output Impedance (Maximum) <10 Ohm (Maximum) / <1 Ohm (Typical)
Current 100mA / channel
Power 12 VDC, 1.5 A
Enclosure Plastic
Dimensions 17.5 cm W, 13.5 cm D, and 6 cm H

Transducers compatible with the IA-400D

Design Your System

Choose the parameters you want to Measure

ECG/EKG
EMG : Number of Channels
Invasive BP : Number of Channels
Non-Invasive BP : Number of Channels
Flow : Number of Channels
Temperature: Number of Channels
Force: Number of Channels
Stimulation: Describe the type