RSM Analog Output Settings

This technote provides information on how to export data from Indus Instruments Rodent Surgical Monitor to use with iWorx data acquisition systems. Within the RSM+ application, the users can go to the settings menu in the bottom right-hand side, go to ‘analog output’ to adjust either the channel assignment or the filtering. The users can export 8 channels if they have the 8 channel analog output cable or 4 channels if they have the 4 channel analog output cable.

Configuring Analog Output and Channel Assignment to iWorx DAQ System

Channel Assignment Drop Menus

  • Channel 1 (choose from list on right)
  • Channel 2 (choose from list on right)
  • Channel 3 (choose from list on right)
  • Channel 4 (choose from list on right)
  • Channel 5 (choose from list on right)
  • Channel 6 (choose from list on right)
  • Channel 7 (choose from list on right)
  • Channel 8 (choose from list on right)

Waveforms Available to Output

  • ECG Lead 1 (LA-RA)
  • ECG Lead II (LL-RA)
  • ECG Lead II QRS
  • ECG Lead II Trigger
  • ECG Lead III (LL-LA)
  • Respiration
  • Pulse Plethysmograph (IR)
  • Pulse Plethysmograph (R)
  • Pressure
  • Core Temperature
  • Pad Temperature
Configuring Analog Output to iWorx DAQ System

Analog Output Filters

The filter setting on the Analog Output is based on the animal model selected under EXPERIMENTAL SETUP – ADVANCED option.

Caution: If the filter settings are unchanged then the animal model is selected make sure that Apply Animal Model Limits to Waveform Filters box is checked and/or make sure that the Auto-Sync filters box under ANALOG OUTPUT – FILTERS is checked (they can also be changed using the drop down arrows next to each setting).

  • The High Pass Filter can be set to Off, 0.2Hz, 0.5Hz, 1Hz, 2Hz, or 5Hz.

  • The Low Pass Filter can be set of Off, 1Hz, 2Hz, 5Hz, 10Hz, 20Hz, 30Hz, 100Hz, 200Hz, 500Hz, or 1000Hz.

    • Note: There is also a 2000Hz hardware filter setting that can be used, but if this is used, make sure to uncheck Auto-Sync.

  • The Notch Filter can be set to Off, 50Hz, or 60Hz (the notch filters are hardware filters).

High Pass Filter – This filter stabilizes the waveform baseline.

Low Pass Filter – This filter removes high frequency noise superimposed on the output signal.

Notch Filter – This filter suppresses cyclic 50Hz or 60Hz AC power line noise that can interfere with the instrumentation.

Another option is to click on the ‘Copy from Signal Configuration’ button to select the filter settings from configuration options selected under Presets. 

Updated on October 22, 2025

Design Your System

Choose the parameters you want to Measure

ECG/EKG
EMG : Number of Channels
Non-Invasive BP : Number of Channels
Flow : Number of Channels
Temperature: Number of Channels
Force: Number of Channels
Stimulation: Describe the type

Recover Password

We will contact you with your login info. 

iWorx Newsletter Signup

By subscribing, you agree with our Privacy Policy and provide consent to receive updates from our company.